Chapter 2. Safety 2 - 9
WARNING
Electric shock may result if this system, including all of its externally
mounted recording and monitoring devices, is not properly grounded.
Never remove the cover from the product. Hazardously high voltage
flows through the product. All internal adjustments and replacements
must be made by a qualified Samsung Medison Customer Service
Department.
Always check the product's casing, cables, cords, and plugs for damage
before using the product. Disconnect and do not use the power source if
the face is cracked, chipped, torn, the housing is damaged, or if the
cable is abraded.
Always disconnect the system from the wall outlet prior to cleaning it.
All patient contact devices, such as probes and ECG leads, must be
removed from the patient prior to the application of a high voltage
defibrillation pulse.
The use of flammable anesthetic gas or oxidizing gases (N
2
O) should be
avoided. Doing so may cause an explosion.
Avoid placing the system where it is likely to be difficult to operate, or
disconnect.
Do not use HF surgical equipment with the system. Any malfunctions in
the HF surgical equipment may result in burns to the patient.
The System must only be connected to a supply mains with protective
earth to avoid risk of electric shock.
CAUTION
The system has been designed for 100-240VAC; you should select the
input voltage of any connected printer and VCR. Prior to connecting a
peripheral power cord, verify that the voltage indicated on the power cord
matches the voltage rating of the peripheral device.
An isolation transformer protects the system from power surges. This
continues to operate when the system is on standby.
Do not immerse the cable in liquids. Cables are not waterproof.
Make sure that the inside of the system is not exposed to or flooded with
liquids. In such cases, fire, electric shock, injury, or damage to the
product may occur.
The auxiliary socket outlets installed on this system are rated 100-
240VAC, with a maximum total load of 150VA. Only use these outlets for
supplying power to equipment that is intended to be part of the
ultrasound system. Do not connect additional multiple-socket outlets or
extension cords to the system.
Do not connect any peripheral devices not listed in this manual to the
auxiliary socket outlets of the system.
Do not touch SIP/SOP and the patient simultaneously. There is a risk of
electric shock from current leakage.