Data Interfaces
Error Message
1   2  3  4  5  6  7  8  9  10  11  12  13  14  15  16  17  18  19  20  21  22
S   t  a  t  *  *  *  *  *  E  r  r  *  *  #  #  *  *  *  *  CR  LF
S   t  a  t  *  *  *  *  *  E  r  r  *  #  #  #  *  *  *  *  CR  LF
*: Space
#: Number (2 or 3 digit error number)
G#  Gross value
N  Net value
T  Application tare memory 1
T2  Application tare memory 2
Diff   Difference from calibration 
value
Targ.   Exact adjustment weight 
value
Nom.   Exact calibration weight 
for SBI protocol output
nRef  Reference sample quantity
pRef  Percentage of reference
wRef  Reference piece weight
Qnt   Result from Counting (piece 
count) and Neutral 
Measurement applications
mDef   Target value for animal 
weighing
x-Net  Animal weighing results
Setp   Target value for 
checkweighing
Diff.W   Absolute difference (e.g., in 
kg) in Checkweighing
Lim   Deviation in % in 
Checkweighing
Max  Upper tolerance for checkw.
Min   Min. tolerance for checkw.
Stat  Status
Classx  Classification
Limx  Class limit
D   Percentage (as loss)
Prc   Percentage (as residue)
Wxx%  Reference percentage weight
Cmpxxx  Component xxx
Cont.T   Contents of the tare memory 
in Net-total Formulation
S-Comp   Total of initial weighings for 
Net-total formulation
PT2  Preset tare 
n  Transaction counter 
*G   Sum of gross weights in 
Totalizing 
*N   Sum of net weights in 
Totalizing  
Ser.no   Serial number of the platform 
or indicator 
  Example:  Output weight value of +1255.7 g
1   2  3  4  5  6  7  8  9  10  11  12  13  14  15  16  17  18  19  20  21  22
G  #  *  *  *  *  +  *  *  *  1  2  5  5  .  7  *  g  *  *  CR  LF
Positions 1-6:  ID code, right-justifi ed with spaces
Position 7:  Plus +, or minus - or space
Position 8:  Space
Positions 9-16:    Weight value with decimal point; leading zeros are output as 
spaces (a comma can also be set instead of a decimal point, 
menu item 7.17).
Position 17:  Space
Positions 18-20: Characters for unit of measure, space or ! sign as a symbol
Position 21:  Carriage return
Position 22:  Line feed
 
h
  If the weight value is output with 10-fold increased resolution, this value is not 
permitted to be printed or saved in a weighing instrument operated in legal 
metrology in the SBI mode. In this case, the unit symbol is not included with output.