Operating Instructions Combics Indicators 25
Setting the Preload
Setup Information Setting the preload when weighing in legal metrology is only possible when the
menu access switch is open.
The “Set Preload” function (menu item 1.9.8) must be allocated to the J key
(see page 137).
3
Once the preload has been set, the J key must be reallocated back to its original
function in the Setup menu, e. g. external calibration/adjustment with default
weights (Setup menu item 1.9).
Procedure
t For scales used in legal metrology: Open the menu access switch.
( t Zero the weighing platform.
t Place the preload weight on the weighing platform.
J t Start the “Set Preload” function.
y After a short period of time the preload will be applied and the indicator will
automatically switch back to weighing mode.
t Re-close the menu access switch.