Operating Instructions Combics Indicators  113
Care and Maintenance
Service
Regular servicing by a Sartorius technician will extend the service life of your 
equipment and ensure its continued weighing accuracy. Sartorius offers its 
customers service contracts with regular maintenance intervals ranging from one 
month to two years. The frequency of the maintenance intervals depends on the 
operating conditions and the operator's tolerance requirements.
Repairs
2
  Disconnect the power supply to the defective equipment immediately (unplug the 
power cord from the mains supply). Repair work must be performed by authorized 
Sartorius service technicians using original spare parts. Repairs performed by 
untrained persons may result in considerable hazards for the user. 
2
  If a cable or cable gland is damaged or defective, replace the cable as a complete 
unit with all its connectors.
2
  Do not open the indicator while it is carrying current. Wait at least 10 seconds after 
disconnecting it from power before beginning to open the equipment. Proper fit-
ting of all surfaces is essential for the IP rating of the housing; for this reason the 
device must be opened and closed by a certified technician.
Cleaning
2
  Disconnect the power supply to the scale before cleaning (unplug the power cord 
from the mains supply).
Cleaning the interior of the weighing platform (CAW.. models only)
t   Remove the load plate. Special care is required for models larger than 
1000 x 1000 mm. 
t   Use compressed air to clean out the interior or use a weak water jet (max. 60°C).
Take special care that no dirt gets into the gap of the overload protection (gap 
between the load sensor and screw-on plate.
Corrosive Environment
t  Remove all traces of corrosive substances from the device on a regular basis.
Cleaning the Indicator
2
  Disconnect the power supply to the indicator (unplug the power cord from the 
mains supply). If necessary, disconnect the data cable.
2
  Make sure that no liquid enters the indicator.
2
  Do not use aggressive cleaning agents (solvents or similar agents).
2
  Do not spray the device with water or blow with compressed air.
  Care and Maintenance