Internal Calibration/Adjustment for CAH Models...
  System Requirements  –   “Internal Calibration” must be set in the Setup menu: setup / wp-1/ param.1 / 
CAL./ADJ. / Cal.Int. 
–   The scale housing has a built-in motorized calibration weight.
  Example:  Internal calibration and then adjustment (weighing parameters: factory settings 
for CAH... models)
  Operation
(  1.)  Zero the scale.
J  2.)   Start calibration 
(e. g., when adjustment 
prompt fl ashes WP).
   cal.Int. is displayed for 
two seconds.
 y   The internal calibration 
weight is loaded 
automatically.
 
   The difference between 
the measured value and 
the true weight of the 
sample will be displayed 
with plus/minus signs.
Internal calibration    A printout will be 
Diff. +        0.2 g     generated if the process 
is cancelled using the 
(  key.
J  3.)   Activate calibration/
adjustment (press the ( 
key to cancel).
   The adjustment weight is 
displayed once adjustment 
is fi nished.
--------------------    A GMP-compliant 
24.11.2010     10:13    printout is generated.
Type    CAH1E-34ED-H
Ser.no.     27163010
Vers.  C2 101.081210
BVers.      01-62-01
-------------------- 
Internal calibration 
Diff. +        0.2 g
Internal  adjustment 
Diff. +        0.0 g
--------------------
24.11.2010     10:15
Name:
--------------------