EasyManuals Logo

Sartorius Combics CAW1P User Manual

Sartorius Combics CAW1P
168 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
26 Operating Instructions Combics Indicators
Clearing the Preload
Setup Information Deleting the preload when weighing in legal metrology is only possible when the
menu access switch is open.
The “Delete Preload” function (menu item 1.9.9) must be allocated to the J
key (see page 137).
3
Once the preload has been deleted, the J key must be reallocated back to its
original function in the Setup menu, e. g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
t For scales used in legal metrology: Open the menu access switch.
t Remove the preload weight from the weighing platform.
) hold t Start the “Delete Preload” function.
y After a short period of time the preload will be deleted and the indicator will
automatically switch back to weighing mode.
t Re-close the menu access switch.
Confi guring Weighing Platforms

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Combics CAW1P and is the answer not in the manual?

Sartorius Combics CAW1P Specifications

General IconGeneral
BrandSartorius
ModelCombics CAW1P
CategoryScales
LanguageEnglish

Related product manuals