Clearing the Preload
Setup Information – Deleting the preload when weighing in legal metrology is only possible when the
menu access switch is open.
– The “Delete Preload” function (menu item 1.9.9) must be allocated to the J
key (see page 137).
3
Once the preload has been deleted, the J key must be reallocated back to its
original function in the Setup menu, e. g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
t For scales used in legal metrology: Open the menu access switch.
t Remove the preload weight from the weighing platform.
) hold t Start the “Delete Preload” function.
y After a short period of time the preload will be deleted and the indicator will
automatically switch back to weighing mode.
t Re-close the menu access switch.