Operating Instructions Combics Indicators 93
Configuring Printouts
Purpose You can individually defi ne each measurement printout. This should be carried out
after setting the applications since some data in the printout is application-
dependent.
In the “Print parameters” menu, single, component and total data records can be
confi gured, which contain the available print items for the respective applications.
Using the total data record for “Totalizing” and “Net-total Formulation”
applications, you can defi ne which parameters are printed using the c key.
Features – Six lists each with a max. length of 30 print items
– Single printout Printer 1
– Component printout Printer 1
– Total data printout Printer 1
– Single printout Printer 2
– Component printout Printer 2
– Total data printout Printer 2
– Single, component and total data records can be confi gured separately
– Print single printout: p key
Auto printout of application when Setup menu is activated:
– Animal weighing (averaging)
– Checkweighing
– Classifi cation
– Print component printout (Combics 2 only):
Totalizing/Net-total formulation with the O key
Setting: Applic./Applic.3/TOTALIZ printout: component printout
– Print totalizing printout (Combics 2 only):
For selected application Totalizing/Net-total formulation with c key
– When switching to another application in Setup, only the application-
dependent printout lists are deleted. The other printout lists remain saved.
– Print items can be deleted individually: Press and hold the ( key
– Print items “Form Feed” for record footer:
Move to the next label start for printer type: YDP14IS: “Label” and YDP04IS,
setting “Label, manual form feed"
– ISO/GLP/GMP-compliant printout: The Setup menu confi guration under “ISO/
GMP-compliant printout” is also active for confi gured printouts.
Preparation t Open Menu mode (see page 31).
k k ... t Select the Setup menu.
k k ...) t Select and open the Print submenu.
k k ...) t Select and open the PROTOC. submenu.
Operation