108 Operating Instructions Combics Indicators
Data Interfaces
Configuring Printouts
Printouts are confi gured in the Setup menu under “Printouts” (setup / print /
protoc.). This should be carried out after confi guring the application since some
data in the printout is application-dependent.
You can confi gure a separate printout for each interface Each printout is comprised
of different information blocks that can be activated or deactivated via multiple
selection in the menu.
Combics 2 only: For the “Totalizing” and “Net-total Formulation” applications,
the totalizing/results printout can be confi gured independent of the
individual/ component printout.
Headers
2 headers each with a max. of 20 characters are available (e.g. for printing the
company name).
Input: menu items 7.4.1 and 7.4.2.
Empty headers are not printed:
Example Print image:
ACE HARDWARE
GOETTINGEN
In this example, the company name is printed centered because there are 4 and 5
spaces before the text.
GMP-compliant printouts
When this function, the printout is supplemented with a GMP header and a GMP
footer (GMP: “Good Manufacturing Practice”).
Setting: Menu item 7.13.
The GMP header precedes the rst measured result. The GMP footer is printed
either after each individual measurement result (“GMP-compliant printout always
for 1 result.” menu item 7.13.2) or after the last result in a series of measurements
(“GMP-compliant printout always for several application results,” menu item
7.13.3). To end a series of measured results, press and hold the p key. In this case,
the p symbol is displayed after the GMP header is printed and remains in the display
until the GMP footer is printed.
If you toggle to a different platform (Combics 2 only) while a GMP printout of
several measured results is being generated (menu item 7.13.3), the GMP footer
for the platform used up to that point is generated when you press the n key.
The GMP header for the other platform is included on the next printout generated.
A GMP-compliant printout is generated automatically at the conclusion of
calibration/adjustment, linearization routines, as well as when you set or clear a
preload.
When printing GMP-compliant printouts on label printers under menu setting
7.13.3, the relationship between the GMP header and footer is lost (printed on
different labels). GMP-compliant printouts on label printers, therefore, should only
take place using menu setting 7.13.2.
Three examples of GMP headers and one example of a footer are shown in the
following. On Combics 1 models, the “date and time” line is not included.