Operating Instructions Combics Indicators 117
CAW*S... Model Resolutions
Resolution 1 weighing range Resolution* 2 weighing ranges
CAW*S... -L -I -BCE -NCE
Weighing Weighing Resolution Weighing Resolution
range 15,000d 30,000 d 1 + 3000e range 1 Range 1 range 2 Range 2
in kg in g in g in g in kg in g in kg in g
3 0.2 0.1 1 1.5 0.5 3 1
6 0.5 0.2 2 3 1 6 2
15 1 0.5 5 6 2 15 5
30 2 1 10 15 5 30 10
60 5 2 20 30 10 60 20
150 10 5 50 60 20 150 50
300 20 10 100 150 50 300 100
600 50 20 200 300 100 600 200
1500 100 50 500 600 200 1500 500
3000 200 100 1000 1500 500 3000 1000
2 + 3000e
1
) 3 + 3000e
1
)
2
) 6000e
1
)
2
)
Multi Intervall -MCE Multi Range
-UCE -RCE
Weighing
Resolution
Weighing
Resolution
Weighing
Resolution
Weighing
Resolution
Weighing
Resolution
Weighing
range 1 range 2 range 1 range 2 range 3 range
Range 1 Range 2 Range 1 Range 2 Range 3
in kg in g in kg in g in kg in g in kg in g in kg in g in kg in g
1.5 0.5 3 1 - - - - - - - -
3 1 6 2 1.5 0.5 3 1 6 2 - -
- - - - 3 1 6 2 15 5 - -
15 5 30 10 6 2 15 5 30 10 - -
30 10 60 20 15 5 30 10 60 20 60 10
60 20 150 50 30 10 60 20 150 50 120 20
150 50 300 100 60 20 150 50 300 100 300 50
300 100 600 200 150 50 300 100 600 200 600 100
600 200 1500 500 300 100 600 200 1500 600 1200 200
1500 500 3000 1000 600 200 1500 500 3000 1000 3000 500
* The weighing ranges permitted for use in legal metrology are listed in the Declaration of Conformity. The respective weighing range cannot be shifted for devices
with two weighing ranges (2 + 3000e). If a higher weighing range is used for weighing, then the lower resolution remains after taring.
1
) IP69K not for all CAW*S4 models (see data sheet)
2
) Depending on the model type (see data sheet)
Resolution (depending on the readability)
L 15,000 d, intervals
I 30,000 d, intervals
BCE 1 + 3000 e, verifi able increments
NCE 2 + 3000 e, verifi able increments, multi-range, ne range only for spec. capacity level
MCE 2 + 3000 e, verifi able increments, multi-interval, ne range can be continuously used
UCE 3 + 3000 e, verifi able increments, multi-range, ne range only for spec. capacity level
RCE 1 + 6000 e, verifi able increments
Specifi cations

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Combics CAW1P and is the answer not in the manual?

Sartorius Combics CAW1P Specifications

General IconGeneral
BrandSartorius
ModelCombics CAW1P
CategoryScales
LanguageEnglish

Related product manuals