Operators Handbook DCU3 – System
10., Main Function „Documentation“
BAEDCU3 - Rev. 4.6 - 0504 10 - 2
10.2.1 Operating Display
Tag Entry Displayed value or function, required input
NAME 1 ... 6 input of the name of the record: 12 selectable characters
ALARM selection of the printing state of the alarm record
enable
− alarms are printed
disable
− alarms are not printed
CYCLE 1 ... 60 input of print sequence of trend record, in minutes
PARAMETER selection of additional records
calib
− calibration data
alarm
− limits of process value alarms
MODE operation mode for record
start
− selected records are printed
stop
− interruption of record printout
TYPE input of parameter for trend record (max. 10 x)
CL SP
− controller setpoint
PV MEAN
− mean process value
PV MIN
− minimum process value
PV MAX
− maximum process value
PV
− actual process values at time of output to printer
10.2.2 Special Notes
You can connect conventional standard dot-matrix printers to print the event record. The
printer is attached to the DCU-System at the peripheral port (RS232). The printer must have
fonts for printing at 17 cpi (characters per inch).