Operators Handbook DCU3 – System 
 
5., Main Function „Calibration“ 
BAEDCU3 - Rev. 4.6 - 0504    5 - 8 
5.6  Totalizer for Pumps and Valves 
For monitoring of the corrective agent consumption the dosing times of the pumps or of the dosing 
valves are totalized and used as process values. The DCU3 - system converts the dosing times into 
delivered volumes considering the specific flowrates of the pumps or of the dosing valves. 
The dosing counters can be zeroed at any time via the operating display. If the specific flowrates are 
known, you can enter them directly via the touch display. If the flowrates must be evaluated you can 
use the „Calibration Totalizer“ function. Furthermore the flowrates can be calculated automatically dur-
ing the calibration process from the measured running time and the entered delivery volume. 
The standard configurations of the DCU-system include calibration routines and dosing counters for 
acid, base, antifoam, feed, harvest and substrate. Since the calibration and dosing counter functions 
are the same for all pumps and dosing valves, only those for the acid pumps will be described herein. 
5.6.1 Operating Display 
 
Tag  Entry  Displayed value or function, required input 
ACID  ml  Display of delivered volume of liquid 
FLOW  ml/min  Display of calculated delivery or input of either specific 
pump delivery or valve flow, as far as known 
Start Calibration    Enabling the calibration 
Reset Totalizer    Resetting the dosing counter to zero 
PV ALARM    Enabling the alarm display 
CLOOP    Selection of the controller menu 
  Additional touch keys, values and functions available in this display:: 
VOLUME  ml  Input of volume delivered during calibration