14., Supplement
BAEDCU3 - Rev. 4.6 - 0504 14 - 5
Text of alarm line Meaning Remedial measures
„2.SEC/not inst./102“ 1
st
serial interface extension MVME335
is missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„VDO32/not inst./102“ Digital output extension VDOT-32 is
missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„VDO32/DOXXX-YYY/ZZZ“ Error of the digital output channels
XXX-YYY or digital output extension
VDOT-32; ZZZ = error code
Check cable connection;
check supply voltages
„VDO32/DIXXX-YYY /XXX“ Error of the digital input channels XXX-
YYY or digital output extension VDOT-
32; ZZZ = error code
Check cable connection;
check supply voltages
„VDO64/not inst./102“ Digital output extension VDOT-64 is
missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„VDO64/DOXXXYYY/ZZZ“ Error of the digital output channels
XXX-YYY of digital output VDOT-64;
ZZZ = error code
Check cable connection;
check supply voltages
14.2 Trouble Shooting for the DCU-System
The Sartorius BBI Systems GmbH develops, manufactures and configures the DCU system with ut-
most care. If problems arise with the use of the DCU system (malfunctions, technical problems), the
user or the customer service should return a detailed description of the problem.
The operator or the customers´ service can use the form „Function bug handling“ which is available on
request, to describe the problem. If all aspects are considered this will enable the Sartorius BBI Sys-
tems GmbH to answer within the shortest possible time offering a solution for the problem.
If problems arise with the DCU system we recommend that the user or the customers´ service should
contact the responsible local service representative of Sartorius BBI Systems GmbH first. However, if
no local service representative is available, he can directly send the completed form to:
Sartorius BBI Systems GmbH
Dept. FA (Product engineering, automation projects)
Schwarzenberger Weg 73-79
D - 34212 Melsungen, Fed. Republic of Germany
phone +49 56 61 - 71 34 00, fax +49 56 61 - 71 37 02
14.3 Interlocking Functions
Interlocking functions are finally configured and cannot be changed by the user. Outputs
controlled and locked by an interlocking function are indicated by the state message „lock“
in the entry for manual access of the corresponding menu.
The extend of interlocking functions is specific for the fermentor and will be recorded in the
configuration tables. These tables are included in the delivery of the DCU-system.