Operators Handbook DCU3 – System 
 
6., Main function „Control Loops“ 
BAEDCU3 - Rev. 4.6 - 0504    6 - 24 
6.15 Foam Controller 
When foam comes into contact with the antifoam sensor, the foam measurement amplifier generates a 
limit value signal for the antifoam controller. This signal is active as long as foam has contact with the 
probe. The sensitivity of the foam detection can be adjusted via the display of the foam controller. 
The output of the antifoam control function in the DCU system triggers a pump or a dosing valve to 
supply antifoam agent to the culture. The pump or dosing valve is switched on and off periodically. 
The dosing times and cycle time for intermittent operation can be entered into the control display. 
6.15.1 Operating Display 
 
Tag  Entry  Displayed value or function, required input 
FOAM  on / off  Display of foam signal („on“ = foam) 
SENSI  low / high  Input of sensitivity of foam signal amplifier  
MODE    Input of operating mode of controller 
 off 
−  controller off, OUT = off 
 auto 
−  controller on 
 manual 
−  manual access to controller output 
OUT  on / off  Controller output („on“ = active); data entry only in „manual“ mode 
CYCLE  m:s  Input of cycle time in [minutes:seconds] 
PULSE  m:s  Input of running time of pump [minutes:seconds] 
DI ALARM    Enabling the display of digital input alarms