Operators Handbook DCU3 – System
14., Supplement
BAEDCU3 - Rev. 4.6 - 0504 14 - 4
Text of alarm line Meaning Remedial measures
„3.FIB/AI01-16 /175“ The analog inputs AIP01...AIP17 of the
3rd FIB are malfunctioning
Send a claim note for FIB
„3.FIB/AI17-32 /175“ The analog inputs AIP17..AIP32 of the
3rd FIB of the analog input extension
are are malfunctioning
Send a claim note for FIB
„3.FIB/AI17-32 /102“ The analog input extension of the 3rd
FIB is missing or installed incorrectly
Check whether analog input
extenision is installed (prop-
erly); if yes, send a claim for
analog input extension
„3.FIB/AO01-04 /175“ The analog outputs AOP01..04 of the
3rd FIB are malfunctioning
Send a claim note for the FIB
„3.FIB/SIO Ser6./XXX“ Error of peripheral interface of the 3rd
FIB; XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„3.FIB/SIO Ser5./XXX“ Error of host interface of the 3rd FIB;
XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„4.FIB/not inst./102“ The 4th FIB is missing or installed in-
correctly
Check board address; enable
(return to) delivery state,
otherwise send a claim for FIB
„4.FIB/AI01-16 /175“ The analog inputs AIP01...AIP17 of the
4th FIB are malfunctioning
Send a claim note for the FIB
„4.FIB/AI17-32 /175“ The analog inputs AIP17..AIP32 of the
4th FIB of the analog input extension
are are malfunctioning
Send a claim note for the FIB
„4.FIB/AI17-32 /102“ The analog input extension of the 4th
FIB is missing or installed incorrectly
Check whether analog input
extenision is installed (prop-
erly); if yes, send a claim for
analog input extension
„4.FIB/AO01-04 /175“ The analog outputs AOP01..04 of the
4th FIB are malfunctioning
Send a claim note for the FIB
„4.FIB/SIO Ser6./XXX“ Error of peripheral interface of the 4th
FIB; XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„4.FIB/SIO Ser7./XXX“ Error of host interface of the 4th FIB;
XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„1.MVM/not inst./102“ 1
st
. serial interface extension MVME335
missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„2.MVM/not inst./102“ 1
st
. serial interface extension MVME335
missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB
„1.SEC/not inst./102“ 1
st
serial interface extension MVME335
is missing or installed incorrectly
Check board address;
reset to delivery state;
send a claim not for FIB