Operators Handbook DCU3 – System
14., Supplement
BAEDCU3 - Rev. 4.6 - 0504 14 - 3
14.1.4 Service and Installation Messages
Text of alarm line Meaning Remedial measures
„1.FIB/not inst./102“ 1
st
FIB missing or installed incorrectly Check board address;
reset to delivery state;
send a claim note for the FIB
„1.FIB/AI01-16 /175“ Analog inputs AIP01..AIP17 of the 1
st
FIB are defective
Send a claim note for the FIB
„1.FIB/AI17-32 /175“ Analog inputs AIP17..AIP32 of the 1
st
FIB or the analog input extension are
defective
Send a claim note for the FIB
„1.FIB/AI17-32 /102“ Analog input extension of the 1
st
FIB is
missing or installed incorrectly
Check whether analog input
extenision is installed (prop-
erly); if yes, send a claim for
analog input extension
„1.FIB/AO01-04/175“ Analog outputs AOP01..04 of the 1
st
FIB are defective
Send a claim note for the FIB
„1.FIB/SIO Host /XXX“ Error on host interface of the 1
st
FIB;
XXX = error code
Check cable connection;
check DCU address
„1.FIB/SIO Ser2./XXX“ Error of the peripheral interface of the
1
st
FIB; XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„2.FIB/not inst./102“ 2nd FIB missing or installed incorrectly Check board address;
reset to delivery state;
send a claim note for the FIB
„2.FIB/AI01-16 /175“ analog inputs AIP01..AIP17 of the 2
nd
FIB are defective
Send a claim note for the FIB
„2.FIB/AI17-32 /175“ Analog inputs AIP17..AIP32 of the 2
nd
FIB of the analog input extension are
defective
Send a claim note for the FIB
„2.FIB/AI17-32 /102“ Analog input extension of the 2
nd
FIB
missing or installed incorrectly
Check whether analog input
extenision is installed (prop-
erly); if yes, send a claim for
analog input extension
„2.FIB/AO01-04 /175“ Analog outputs AOP01..04 of the 2
nd
FIB are defective
Send a claim note for the FIB
„2.FIB/SIO Ser4./XXX“ Error of the peripheral interface of the
2nd FIB; XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„2.FIB/SIO Ser2./XXX“ Error of the host interface of the 2
nd
FIB; XXX = error code
Check cable connection;
check peripheral device (may
be malfunctioning)
„3.FIB/not inst./102“ 3rd FIB missing or installed incorrectly Check board address; enable
(return to) delivery state,
otherwise send a claim for FIB