Operators Handbook DCU3 – System
5., Main Function „Calibration“
BAEDCU3 - Rev. 4.6 - 0504 5 - 12
5.8.2 Special Notes
1. It is recommende to set a cycle interval of 10 minutes.
2. When selecting an interval of 10 or 60 minutes, respectively, the calculation and display of
the flowrate starts with the shortest interval and automatically switches over to the next
longer one. The flowrate of the initial phase is calculated from the shortest interval. This is
also true when a new setpoint is defined for the dosing controller.
The same is true for presetting a new setpoint of the dosing controller.
3. For manual refilling switch over to „Refill“-mode.
The automatic calculation of the actual flow will be stopped. The dosing controller continues
to act upon the last flow calculated. You can now refill the substrate storage vessel (bottle)
and tare the balance again. After switching over to „contr.“ mode the system starts with re-
calculation of the flow and continues dosing, now considering the actual values.