Operators Handbook DCU3 – System
7., Main function „Batch Control“
BAEDCU3 - Rev. 4.6 - 0504 7 - 3
7.2.2 Operating Display „Automatic Sequential Control“
Note: This example is only valid for BIOSTAT
®
ED:
Tag Entry Displayed value or function, required input
STATE Enter the sterilization program mode „start“ or „stop“
start
start sterilization
stop
stop sterilization
step
not allowed (only valid for single step controls)
Display of the status of the sterilization program:
active
sterilization program running
lock
start is impossible since another sequence or recipe is active
inactive
sterilization program inactive
ACK Request for „Acknowledgement-display“, see „alarm messages“ in
the supplement
PHASE Display of the sterilization step
----
sterilization program inactive
HEAT
heating to sterilization temperature
STERI
control of sterilization temperature
COOL1
cooling to 80 degC
COOL2
cooling from 80 degC to process temperature
READY
process temperature reached (message: „sterilization finished“)
TEMP Display of the current fermentor temperature
ELAPS Display of elapsed sterilization time in the sterilization phase
STEMP Input of sterilization temperature
FTEMP Input of fermentor (process) temperature
STIME Input of sterilization time