Overview
Specifications
Standard Models
General Specifications:
AC power source/power requirements AC adapter, 230 or 115 V, +15% ... – 20%
Frequency 48 – 60 Hz
Allowable ambient operating temperature 0 ... +40 °C (273 ... 313 K, 32 °F ... 104 °F)
Operating temperature range + 10 ... + 30 °C
Dust and water protection rating according
to EN 60529 IP54
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update 0.1 – 0.4
(depends on the filter level selected)
Power consumption 16 VA: maximum; 9 VA: average
Hours of operation with fully charged
YRB 06 Z external battery pack, approx. 14 h
Selectable weight units Grams, kilograms, carats, pounds, ounces, Troy ounces, Hong Kong taels, Singapore taels,
Taiwanese taels, grains, pennyweights, milligrams, parts per pound, Chinese taels, Mommes,
Austrian carats, Tola, Baht and Mesghal
Selectable application programs Mass unit conversion, counting, weighing in percent, animal weighing, calculation, over/under
checkweighing, time-controlled functions, totalizing, statistics, 2nd tare memory, identifiers
Built-in interface RS-232C
Format: 7-bit ASCII, 1 start bit,
1 or 2 stop bits
Parity: odd, even or space
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
Model-Specific Specifications:
Model FC06BBE-S FC6CCE-H FC2CCE-S FC12CCE-S FC6CCE-S
Readability g 0.001 0.01 0.01 0.1 0.1
Weighing capacity g 620 6,200 2,200 12,000 6,200
Max. capacity kg 3 25 10 50 50
Tare range (by subtraction) g – 620 – 6,200 – 2,200 – 12,000 – 6,200
Electronically compensated preload g 93 110 1,200 1,240
(without restricting weighing range)
Max. preload when starting g 110 5,200 1,300 10,000 6,400
calibration/adjustment (isoTEST)
(scale must be zeroed)
Reproducibility (standard deviation) ≤±g 0.001 0.01 0.01 0.05 0.05
Linearity ≤±g 0.002 0.02 0.02 0.2 0.1
Sensitivity drift within +10 ... +30 °C ≤±/K 2 • 10
-6
2 • 10
-6
2 • 10
-6
4 • 10
-6
4 • 10
-6
Response time (average) s 1.5 1.5 1.5 1 1
External calibration weight g 500 (E2) 5,000 (E2) 2,000 (F1) 5,000 (F1) 5,000 (F2)
(of at least accuracy class...)
Pan size mm Ø 130 218 x 200 218 x 200 218 x 200 218 x 200
Dimensions (W x D x H) mm 240x294x86 240x294x86 240x294x86 240x294x86 240x294x86
Net weight, approx. kg 7 8.4 7.3 7.3 7.3
Dust and water protection rating
according to EN 60529 IP54