Page #53 background imageLoading...
Page #53 background image
Operating the Scale
53
Animal Weighing C
Purpose
Use this program to determine the
weights of unstable samples (e.g.,
live animals) or to determine weights
under unstable ambient conditions.
In this program, the scale calculates
the weight as the average of a
defined number of individual
weighing operations. These
weighing operations are also known
as “subweighing operations.”
You can use this application program
in combination with a program
chosen from Application 2
(checkweighing, time-controlled
functions) and one from Application
3 (totalizing, net-total formulation,
statistics).
Available Features
Animal weighing started manually
or automatically
Optional scale configuration in
the Setup menu for automatically
initializing this application
when you turn on the scale
Minimum load threshold
(100 display increments) for auto-
matic start of animal weighing
Start range: Automatic start of
animal weighing operation when
three successive subweights lie
within a user-defined tolerance
range (calm = 2%, normal = 5%,
active = 10%)
Number of weighing operations
for calculation of an average
(mDef) can be set before
the beginning of each animal
weighing operation
The factor for calculation
of the result can be set before
the beginning of each animal
weighing operation
The number of subweighs
remaining to be performed is
indicated in the text display
during weighing
Arithmetic average displayed as
a result in the pre-set weight unit
(identified by the a symbol).
Optional multiplication of the
arithmetic average by a user-
defined factor Mul.
A circle “o” is displayed as
weight unit and Mul = xxx
is shown in the text line
Toggling between the weight
and the calculated value by
pressing the xNet soft key and
the xRes soft key
Automatic output of results via the
interface port:
Number of weighing
operations mDef
Multiplication factor Mul
Automatic output of results
(printout) via the interface port:
Weighing result xNet
Calculated result xRes
The unload threshold hold is
equal to one-half the minimum
scale capacity
Return to weighing mode
by unloading the scale; i.e.,
when the load is below the
unload threshold
Factory Settings
Animal activity:
Normal
Start: Automatic
Decimal places in result display:
2 decimal places
Printout:
Average weight only
Soft Key Functions
Restar Automatic start:
Unload scale and
weigh next animal,
if desired
Press key to start
next subweigh
Manual start:
Start next subweigh
mDef Store user-input number of
subweighs for averaging
Mul Store user-input factor as
multiplication factor for
calculated the arithmetic
mean
xNet Toggle to the Animal
weighing application
xRes Toggle to the calculated
animal weighing result

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Factory Series FC06BBE-S and is the answer not in the manual?

Sartorius Factory Series FC06BBE-S Specifications

General IconGeneral
BrandSartorius
ModelFactory Series FC06BBE-S
CategoryScales
LanguageEnglish

Related product manuals