119
Overview
Verified Models with EU Type Approval
General Specifications
AC power source/power requirements AC adapter, 230 or 115 V, +15% ... –20%
Frequency 48 – 60 Hz
Operating temperature range +10 ... +30 °C (+50 ... +86 °F)
Adaptation to ambient conditions By selection of 1 of 4 optimized filter levels
Display update 0.1 – 0.4
(depends on the filter level selected)
Power consumption 16 VA: maximum; 9 VA: average
Hours of operation with fully charged
YRB 06 Z external battery pack, approx. 14 h
Selectable weight units Grams, kilograms
Selectable application programs Mass unit conversion, counting, weighing in percent, animal weighing, calculation, over/under
checkweighing, time-controlled functions, totalizing, statistics, 2nd tare memory, identifiers
Built-in interface RS-232C
Format: 7-bit ASCII, 1 start bit,
1 or 2 stop bits
Parity: odd, even or space
Transmission rates: 150 to 19,200 baud
Handshake: Software or hardware
Model-Specific Specifications:
Model FC06BBE-S0CE FC6CCE-H0CE FC2CCE-S0CE FC12CCE-S0CE FC6CCE-S0CE
Type BA BF 500 MA BF 200 BA BF 500 BA BF 500 BA BF 500
Accuracy class * KKKK K
Increment d* g 0.001 0.01 0.01 0.1 0.1
Max. weighing capacity* g 620 6,200 2,200 12,000 6,200
Verification interval e* g 0.01 0.1 0.1 1 1
Min. capacity* g 0.02 0.5 0.5 5 5
Max. capacity kg 3 25 10 50 50
Tare range (by subtraction) ≤ 100% of the max. weighing capacity
Electronically compensated preload g 93 – 110 1,200 1,240
(without restricting weighing range)
Max. preload when starting g 110 5,200 1,300 8,200 2,440
calibration/adjustment (isoTEST)
(scale must be zeroed)
Application range according to CD* g 0.02 – 620 0.5 – 6,200 0.5 – 2,200 5 – 12,000 5 – 6,200
Response time (average) s 1.5 1.5 1.5 1 1
Pan size mm Ø 130 218 x 200 218 x 200 218 x 200 218 x 200
Dimensions (W x D x H) mm 240x294x86 240x294x86 240x294x86 240x294x86 240x294x86
Net weight. approx. kg 7 8.4 7.3 7.3 7.3
1) = With the isoCAL function deactivated, the verified scale can only be used within the limited temperature range (can only be modified by the
Sartorius Service Center): For scales of accuracy class k: +15°C to +25°C
For scales of accuracy class K: +10°C to +30°C
* CD = Council Directive 90/384/EEC for non-automatic weighing instruments used within the European Economic Area