70
Practical Example
Totalize counted pieces
Settings (changes in the factory settings required for this example):
Setup: App: Application 1: Counting
Setup: App: Application 3: Totalizing: Evaluated values: Net + calculated
Setup: App: Application 3: Totalizing: Evaluation mode, MR function: Final evaluation, display + print
Step Key (or instruction) Display/Output
1. Turn on the scale and configure e/[ON/OFF]
the settings as indicated above
2. Delete old totalization data, c
if necessary
3. Tare the scale t
4. Toggle to Application 1: Counting D/[TOGGLE]
5. Place the displayed number of Place parts to be counted
parts on the scale (here: 10 pcs) on the scale
6. Initialize the Counting application Start soft key
7. Remove the reference sample Unload the scale
quantity and toggle to Totalizing D/[TOGGLE]
8. Place a number of parts on the Place parts on the scale
scale (here: 50 pcs)
nRef 10 pcs
wRef 35.28900 g
isoTST
6
isoTST
6
isoTST
6
isoTST
6
isoTST
6
isoTST
6