EasyManuals Logo

Sartorius Factory Series FC06BBE-S User Manual

Sartorius Factory Series FC06BBE-S
132 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #88 background imageLoading...
Page #88 background image
88
Using Applications in Combination
Example 1: Checkweighing with statistical evaluation
You want to check a piece count, and have the results that lie within the tolerance range statistically evaluated and
printed as a ISO/GMP-compliant record.
Settings (changes in the factory settings required for this example):
Setup: App: Application 3: Statistics: Automatic storage: On, first value at stability
Setup: App: Application 3: Statistics: Source of data for auto storage: Application 2
Setup: App: Application 3: Statistics: Evaluated value: Calculated
Setup: App: Application 3: Statistics: Evaluation mode, MR function: Intermediate evaluation, display+print
Setup: App: Basic application: Printout configuration: ISO/GLP/GMP printout: Always
Step Key (or instruction) Display/Output
1. If necessary: turn on the scale e/[ON/OFF]
and enter the settings given above
2. Place reference sample quantity Place parts on the scale
on the scale
3. Initialize the scale Start soft key
4. Remove reference sample quantity Unload the scale
isoTST
6
isoTST
6
isoTST
6
isoTST
6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Factory Series FC06BBE-S and is the answer not in the manual?