6β36β2
Specifications: Master Series
Model AC 211S-00MS AC 211P-00MS AC 121S-00MS
Weighing range structure SuperRange PolyRange SuperRange
Capacity/weighing range levels g 210 60/120/210 120
Readability g 0.0001 0.0001/0.0002/0.0005 0.0001
Tare range (by subtraction) g β210 β210 β120
Reproducibility (standard deviation) g β€ Β±0.0001 β€ Β±0.0001/0.0002/0.0005 β€ Β±0.0001
Linearity g β€ Β± 0.0002 β€ Β±0.0002/0.0004/0.0005 β€ Β±0.0002
Response time (average) s βββββββββββββββββββββββββββββββββββββββββ β€2.5 ββββββββββββββββββββββββββββββββββββββββ
Shortest response time in the IQ-mode s ββββββββββββββββββββββββββββββββββββββββββ 1.5 βββββββββββββββββββββββββββββββββββββββββ
Adaptation to ambient conditions and application requirements ββββββββββββββββββββββββββ by selection of 1 of 4 optimized filter levels βββββββββββββββββββββββββ
Display update (depends on the filter level selected) s βββββββββββββββββββββββββββββββββββββββ 0.1β 0.4 βββββββββββββββββββββββββββββββββββββββ
Sensitivity drift within 283β303 K /K βββββββββββββββββββββββββββββββββββββββ β€ Β± 2Β·10
β6
βββββββββββββββββββββββββββββββββββββββ
Ambient temperature range K ββββββββββββββββββββββββ 283β303 (+10Β°Cβ¦+30Β°C; +50Β° Fβ¦+86Β° F) βββββββββββββββββββββββ
Pan size mm/in βββββββββββββββββββββββββββββββββββββ 90 Γ/3.5 Γ βββββββββββββββββββββββββββββββββββββ
Weighing chamber (WxDxH) mm/in ββββββββββββββββββββββββββββββ 200x180 x260/8x7.1x10.2 ββββββββββββββββββββββββββββββ
Clearance above the weighing pan mm/in ββββββββββββββββββββββββββββββββββββββββ 254/10ββββββββββββββββββββββββββββββββββββββββ
Net weight, approx. kg/lb βββββββββββββββββββββββββββββββββββββββ 7.5/16.5 ββββββββββββββββββββββββββββββββββββββ
AC power source/power requirements V~ βββββββββββββββββββββββββ via AC adapter, 230 or 115, +15%β¦β20% ββββββββββββββββββββββββ
Frequency Hz ββββββββββββββββββββββββββββββββββββββββ 48β 60 ββββββββββββββββββββββββββββββββββββββββ
Power consumption VA ββββββββββββββββββββββββββββββββββ max.:16; average: 8 ββββββββββββββββββββββββββββββββββ
Selectable weight units βββββββββββββ g, kg, ct, lb, oz, ozt, tlh, tls, tlt, gr, dwt, mg, /lb, tlc, mom, K, tol and bat ββββββββββββ
Application programs formulation; statistics; averaging; real-time clock with timer function for time-controlled measurements/
weighing; calculation of weights; totalling of weights; density determination; ID no. for
βββ
identification of data; mass unit conversion by toggling; tare memory; net total; weighing in percent;
βββ
over/under checkweighing; counting
Automatic zero tracking function (can be turned off by menu code) ββββββββββββββββββββββββββββββββββββ standard feature ββββββββββββββββββββββββββββββββββββ
Built-in interface ββββββββββββββββββ RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
ββββββββ
transmission rate: 150β¦19,200 baud;1 or 2 stop bits; software/hardware handshake
ββββββββ
Standard features/equipment supplied
Dust cover x x x
Analytical draft shield chamber x x x
AC adapter x x x
Hanger for below-balance weighing x x x
βisoCALβ self-calibration function x x x
ISO/GLP-compliant record/printout x x x
Numeric keypad x x x
β²
β²β²β²β²β²β²β²β²β²β²β²β²β² β² β²
β²β²
β²β²β²β²β²β²
β²β²β²β²β²β²β²β²β²β²