Specifications: Master Series
Model AC 211S-00MS AC 211P-00MS AC 121S-00MS
Weighing range structure SuperRange PolyRange SuperRange
Capacity/weighing range levels g 210 60/120/210 120
Readability g 0.0001 0.0001/0.0002/0.0005 0.0001
Tare range (by subtraction) g –210 –210 –120
Reproducibility (standard deviation) g ±0.0001 ±0.0001/0.0002/0.0005 ±0.0001
Linearity g ± 0.0002 ±0.0002/0.0004/0.0005 ±0.0002
Response time (average) s ––––––––––––––––––––––––––––––––––––––––– 2.5 ––––––––––––––––––––––––––––––––––––––––
Shortest response time in the IQ-mode s –––––––––––––––––––––––––––––––––––––––––– 1.5 –––––––––––––––––––––––––––––––––––––––––
Adaptation to ambient conditions and application requirements –––––––––––––––––––––––––– by selection of 1 of 4 optimized filter levels –––––––––––––––––––––––––
Display update (depends on the filter level selected) s ––––––––––––––––––––––––––––––––––––––– 0.1– 0.4 –––––––––––––––––––––––––––––––––––––––
Sensitivity drift within 283–303 K /K ––––––––––––––––––––––––––––––––––––––– ± 2·10
–6
–––––––––––––––––––––––––––––––––––––––
Ambient temperature range K –––––––––––––––––––––––– 283303 (+10°C…+30°C; +50° F…+86° F) –––––––––––––––––––––––
Pan size mm/in ––––––––––––––––––––––––––––––––––––– 90 Ø/3.5 Ø –––––––––––––––––––––––––––––––––––––
Weighing chamber (WxDxH) mm/in –––––––––––––––––––––––––––––– 200x180 x260/8x7.1x10.2 ––––––––––––––––––––––––––––––
Clearance above the weighing pan mm/in –––––––––––––––––––––––––––––––––––––––– 254/10––––––––––––––––––––––––––––––––––––––––
Net weight, approx. kg/lb ––––––––––––––––––––––––––––––––––––––– 7.5/16.5 ––––––––––––––––––––––––––––––––––––––
AC power source/power requirements V~ ––––––––––––––––––––––––– via AC adapter, 230 or 115, +15%…–20% ––––––––––––––––––––––––
Frequency Hz –––––––––––––––––––––––––––––––––––––––– 48 60 ––––––––––––––––––––––––––––––––––––––––
Power consumption VA –––––––––––––––––––––––––––––––––– max.:16; average: 8 ––––––––––––––––––––––––––––––––––
Selectable weight units ––––––––––––– g, kg, ct, lb, oz, ozt, tlh, tls, tlt, gr, dwt, mg, /lb, tlc, mom, K, tol and bat ––––––––––––
Application programs formulation; statistics; averaging; real-time clock with timer function for time-controlled measurements/
weighing; calculation of weights; totalling of weights; density determination; ID no. for
–––
identification of data; mass unit conversion by toggling; tare memory; net total; weighing in percent;
–––
over/under checkweighing; counting
Automatic zero tracking function (can be turned off by menu code) –––––––––––––––––––––––––––––––––––– standard feature ––––––––––––––––––––––––––––––––––––
Built-in interface –––––––––––––––––– RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
––––––––
transmission rate: 150…19,200 baud;1 or 2 stop bits; software/hardware handshake
––––––––
Standard features/equipment supplied
Dust cover x x x
Analytical draft shield chamber x x x
AC adapter x x x
Hanger for below-balance weighing x x x
“isoCAL” self-calibration function x x x
ISO/GLP-compliant record/printout x x x
Numeric keypad x x x
▲▲▲▲▲▲▲▲▲▲▲▲
▲▲