EasyManuals Logo

Sartorius Master series User Manual

Sartorius Master series
122 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
6564
Specifications: Product Line
1
)
Model AC 211S AC 211P AC 121S
Weighing range structure SuperRange PolyRange SuperRange
Capacity/weighing range levels g 210 60/120/210 120
Readability g 0.0001 0.0001/0.0002/0.0005 0.0001
Tare range (by subtraction) g –210 –210 –120
Reproducibility (standard deviation) g ±0.0001 ±0.0001/0.0002/0.0005 ±0.0001
Linearity g ± 0.0002 ±0.0002/0.0004/0.0005 ±0.0002
Response time (average) s ––––––––––––––––––––––––––––––––––––––––– 2.5 ––––––––––––––––––––––––––––––––––––––––
Shortest response time in the IQ-mode s –––––––––––––––––––––––––––––––––––––––––– 1.5 –––––––––––––––––––––––––––––––––––––––––
Adaptation to ambient conditions and application requirements –––––––––––––––––––––––––– by selection of 1 of 4 optimized filter levels –––––––––––––––––––––––––
Display update (depends on the filter level selected) s ––––––––––––––––––––––––––––––––––––––– 0.1– 0.4 –––––––––––––––––––––––––––––––––––––––
Sensitivity drift within 283–303 K /K ––––––––––––––––––––––––––––––––––––––– ± 2·10
–6
–––––––––––––––––––––––––––––––––––––––
Ambient temperature range K –––––––––––––––––––––––– 283303 (+10 °C…+30°C; +50°F…+86° F) –––––––––––––––––––––––
Pan size mm/in ––––––––––––––––––––––––––––––––––––– 90 Ø/3.5 Ø –––––––––––––––––––––––––––––––––––––
Weighing chamber (WxDxH) mm/in –––––––––––––––––––––––––––––– 200x180 x 260/8x7.1x10.2 ––––––––––––––––––––––––––––––
Clearance above the weighing pan mm/in –––––––––––––––––––––––––––––––––––––––– 254/10––––––––––––––––––––––––––––––––––––––––
Net weight, approx. kg/lb ––––––––––––––––––––––––––––––––––––––– 7.5/16.5 ––––––––––––––––––––––––––––––––––––––
AC power source/power requirements V~ ––––––––––––––––––––––––– via AC adapter, 230 or 115, +15%…–20% ––––––––––––––––––––––––
Frequency Hz –––––––––––––––––––––––––––––––––––––––– 48 60 ––––––––––––––––––––––––––––––––––––––––
Power consumption VA –––––––––––––––––––––––––––––––––– max.:16; average: 8 ––––––––––––––––––––––––––––––––––
Selectable weight units ––––––––––––––––––––––– g, kg, ct, lb, oz, ozt, tlh, tls, tlt, gr, dwt and mg ––––––––––––––––––––––
Application programs
––––––––––––
mass unit conversion by toggling; tare memory; net total; weighing in percent;
–––––––––––
over/under checkweighing (classification and filling); counting
Automatic zero tracking function (can be turned off by menu code) –––––––––––––––––––––––––––––––––––– standard feature ––––––––––––––––––––––––––––––––––––
Built-in interface –––––––––––––––––– RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
––––––––
transmission rate: 150…19,200 baud;1 or 2 stop bits; software/hardware handshake
––––––––
Standard features/equipment supplied
Dust cover x x x
Analytical draft shield chamber x x x
AC adapter x x x
Hanger for below-balance weighing x x x
Built-in, motorized calibration weight x x x
1
) Balances of the MC1 Product Line cannot be upgraded to Master series balances but can be retrofitted with an IAC
(Integrated Applications Computer) for complex applications.

Table of Contents