Control Commands with Lower-Case Letters
All functions that can be selected by pressing the appropriate keys on the balance
can also be activated by commands.
wfunction key
ffunction key
Ffunction key
lfunction key
rfunction key
iinfo function key
cclear function key
Perform calibration test
Print/output
model number
Print/output
serial number
Numeric input
x = Any number which may include a plus or minus sign;
can have anywhere from 1to 8 places
. = Decimal point “.” or “,”
_
= Underline (ASCII = 95)
The numeric value may not have more than 8 digits (excluding the decimal point).
Each control command with lower-case letters (such as f, s, t, w and x)
must be terminated by an underline (ASCII = 95 Hex).
ESCtxxxxxxxx
_
CR LF
......
ESC x 2
_
CR LF
ESC x 1
_
CR LF
ESC x 0
_
CR LF
ESC s 3
_
CR LF
ESC s 0
_
CR LF
ESC f 6
_
CR LF
ESC f 5
_
CR LF
ESC f 2
_
CR LF
ESC f 1
_
CR LF
ESC f 0
_
CR LF
4–9