EasyManua.ls Logo

Sartorius MC 210 S - Data Printout;Record for Application Programs; Application Program Data Printout;Record Format

Sartorius MC 210 S
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Data Printout/Record for Application Programs (menu code 8 10 3)
For application programs, the reference data (parameters) can be included
in the printout/record.
Automatic output of the reference data (parameters) Code
Off 7 1 1*
Reference %/qty. and reference weight 7 1 2
Reference weight only 7 1 3
You can also add the reference data to the printout/record later by pressing i
(information function), p and then either the f or F key.
To have data printed out, perform the following:
– Press the f or F key to output the printout heading and reference data
(the reference data will be stored at the same time)
or
– Press the p key to output the printout heading and the first value
If you input and store new references while an ISO/GLP-compliant record is being
printed out, the new reference data will be output. If you enter different data
before generation of an ISO/GLP-compliant record is started, the printout heading
and the reference data will automatically be printed once you press p. Then
the measured value will be output.
– Press p to output weighing data
– Press the c key to end printout generation (generation of an ISO/GLP-compliant
printout also ends once a calibration or linearization operation has been started)
– Then press the c key to clear the reference data for the application programs
During data printout, the print symbol P appears on the display of the balance.
* = factory setting
2–28

Table of Contents