EasyManua.ls Logo

Scientifica MDU - No high-voltage reading on either detector regardless of enable switch position or voltage controls.; No high-voltage reading on one detector whilst the second detector is fully functional.

Default Icon
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 45 of 58
o You are strongly advised to set both sensitivity / high voltage controls to no more than 20% of
their travel and to keep the light level relatively low during these checks. This will be sufficient
to show correct operation without damaging the detectors.
o Remove each detector cable from the rear of the controller, and inspect the plugs for bent or
missing pins: there should be 8 pins visible in each mini-DIN plug. Re-connect the plugs into
the controller and ensure that each one is pushed fully home; there should be no significant
resistance when the plugs are inserted into their sockets.
o Check both the detector cables for damage.
o Open the filter cube door and remove the filter cube. Check the O-ring around the door to see
that it is properly seated in its groove, and look at the top left corner of the opening to locate the
door-sensing micro-switch. There should be no debris or dirt around the micro-switch, and no
damage to the small actuating arm. Re-close the door and screw it shut securely until there is
no gap between the door and the optical block body. Re-check to see if correct function is
restored.
o Cycle the lever control on the front of the optical block to move the laser / visible dichroic in and
out of the beam path. The motion should be free and un-impeded. Move the lever to the right
(fully clockwise). This should de-activate the interlock. Check to see if correct function is
restored.
o If none of these checks are successful in restoring operation, try the checks in the following
section, but apply them to both detector modules.
No high-voltage reading on one detector whilst the second detector is fully functional.
Probable cause: faulty interlock signal contacts on one detector module.
o As for the previous tests, check the cable and the connector for the offending detector module
and verify that there has been no damage.
o Turn off the controller and un-plug both detectors from it.
o Follow the procedure in the appendix for removing the offending detector module.
o Inspect the spring contacts in the small slot in front of the detector module: they should be un-
distorted, clean and clear of debris / dirt and positioned along the centre of the slot.
o Inspect the contact pins which emerge from the optical block. These should be clean and free
of any debris. They can be cleaned using a lint-free tissue or cotton swab soaked in a little
isopropyl alcohol or industrial methylated spirit. Do not use a more aggressive solvent (no
acetone or chlorinated solvents).
o Check that the filter cube door is fully closed, and that the visible / infrared dichroic mirror is in
the beam (actuator moved fully clockwise). Using a multi-meter set for resistance or continuity
measurement check the resistance between the outer contacts on the main optical block: there
should be a very low resistance reading (sub 1 Ohm).
o Follow the procedure in the appendix for replacing the offending detector module, paying
special attention to the alignment of the module with respect to the top of the optical block.
o Check to see that correct function is restored. If you have no success, the fault may lie
elsewhere: contact Scientifica for assistance.

Table of Contents

Related product manuals