Fit the PMT module(s) so that they are level and flush with the top surface of the optical block.
Before tightening the two long M3 bolts, look at the top surface of the optical block and the detector
module(s): the detector module should be level so that its top surface is flush with that of the optical
block. If this is not the case, then very gently twist the block around the lens holder until the module is
level. Finally, tighten both long M3 bolts.
10.0 Warranty, Technical Queries and Returns
The standard warranty for all Scientifica designed and manufactured goods are two-years. However,
Scientifica’s Multiphoton Imaging system, includes components from other companies, which offer a twelve-
month warranty.
For an extended warranty on the full system (including external companies) please contact your Scientifica
representative. All warranties cover defects in manufacturing and materials. In this unlikely event, Scientifica
will manage the repair and replacement of all components.
The warranty will not apply if the instrument has been damaged by accident, misuse, or as a result of
modification by persons other than Scientifica Ltd personnel.
Before returning an instrument, please obtain a returns number from your local representative or Scientifica
Ltd in the UK.
Please refer to our terms and conditions of sales for information on our legal obligations. However, we pride
ourselves on going beyond the call of duty and seek to remedy where we can and in a timely manner.
For any warranty queries please contact your local distributor or Scientifica Ltd directly.