Operating connected seca measuring devices • 69
17-10-01-266-002a/07-2018B
6.8 Performing final work
Once you have completed the retrofitting of the measuring devices, perform
the following steps:
► Ensure that the device is positioned on the ground flat and stable.
► Ensure that no cables or other parts touch the weighing platform.
► Perform a function check of the device as described in the respective
instructions for use.
► Remove all tools and materials.
► Clean the floor and other installation areas.
► Clean and disinfect the measuring devices as described in the respective
instructions for use.
► Clean and disinfect your hands.
You have the following options to continue:
► Add a seca measuring device in the
secaconnect 103 software:
➔ Adding a seca measuring device
► Change the configuration of a seca measuring device in the
secaconnect 103 software: ➔ Changing a setting of a seca measuring
device
7. OPERATING CONNECTED SECA MEASURING DEVICES
Functional limitations If you connect seca measuring devices to the secaconnect 103, a number of
functions of the seca measuring devices may not be used or only used with
limitations. Otherwise, the following faults can occur:
• Malfunction of individual devices
• Transmission of invalid measured values
• Incorrect allocation of measured values to patient files
► Observe the instructions for use of the seca measuring devices.
► Inform your users about these limitations. To do so, use the quick
reference “Measurement procedure” ➔ Annex: Quick reference:
Measurement procedure.
➔ Functional limitations
➔ Workflow for measurement procedure
Function Device type How to use
hold All Do not use
bmi
Column scale
Measuring stations
Multi-functional scales
Do not use
send/print
a
All seca 360° devices Do not use
2 in 1
Column scale
Measuring stations
• Press the button after each
measurement to delete 2 in 1
value from device memory
• Deactivate ultrasound length
measurement (for example,
models
seca 287, seca 286)
clear Measuring stations
Press button
before each
measurement to delete old
length measurement value from
device memory