Operation • 29
17-10-07-652-100/05-2019B
5.3 Measuring weight
Weighing CAUTION!
Injury from incorrect position of the measuring flap
Injuries may result if the measuring flap is at patient height when
folded up.
► Ensure that the measuring flap is above patient height before the
patient steps onto the weighing platform.
CAUTION!
Incorrect measurement as a result of force shunt
If the patient is in contact with parts other than the weighing platform
during the weight measurement, measuring results will be incorrect.
► Ensure that the patient is in contact only with the weighing platform
during weight measurement.
1. Switch on the measuring station (➔ “Switching on the device”, page 28).
2. Ask the patient to step onto the weighing platform:
– Do not touch the measuring rod
– Do not touch the column and the display head
3. Read off the measuring result.
Taring off additional weight (Tare) Use the Tare function to prevent an additional weight (e.g. a walking stick)
affecting the patient’s weight value.
CAUTION!
Incorrect measurement as a result of force shunt
If an additional weight is in contact with the surface on which the
device is resting, measuring results will be incorrect.
► Make sure that additional weights are placed only on the weighing
platform.
NOTE
The maximum weight which can be displayed is reduced by the
weight of the objects which have been tared off.
1. Switch on the measuring station (➔ “Switching on the device”, page 28).
2. Place the additional weight on the weighing platform.
3. Press the
Tare key.
The message NET is displayed.
4. Wait until the display unit stops flashing and
0.0 is displayed instead.
NOTE
If you remove the additional weight again, - - - - - will be displayed.
5. Ask the patient to step onto the weighing platform with the additional
weight (or carrying the additional weight).
The patient's weight is displayed. The additional weight is deducted
automatically.
6. Read off the measuring result.
7. To deactivate the
Tare function, press the Tare key.
NOTE
When you switch off the measuring station, the Tare function is
deactivated automatically.