38 •
17-10-07-652-100/05-2019B
6.3 Sterilizing
This device may not be sterilised.
7. FUNCTION CHECK
Perform a function check prior to each use.
A complete function check includes:
Visual inspection for mechanical damage
Checking the alignment of the device
Visual and function check of the display elements
Function check of all the controls shown in the section entitled
“Overview”
Function check of optional accessories
If you notice any faults or deviations during the function check, first try to
resolve the error with the aid of the section entitled “Troubleshooting” in this
document.
CAUTION!
Personal injury
If you notice any faults or deviations during the function check which
cannot be resolved with the aid of the section entitled “Troubleshoot-
ing” in this document, you may not use the device.
Have the device repaired by seca Service or by an authorized
service partner.
Follow the section entitled “Servicing” in this document.
8. TROUBLESHOOTING
8.1 General faults
Interval Component
Before
every
measurement
Weighing platform
Head slide with measuring flap
After
every
measurement
Weighing platform
Head slide with measuring flap
As required
Column and display head of measuring
station
Telescopic elements of the measuring rod
Recessed handle
Problem Cause Remedy
No weight display with a load
The device has no power supply
Switch on the measuring station
In the case of battery operation: Insert
batteries
In the case of power supply operation:
Connect the plug-in power supply unit
0.0 does not appear before
weighing
A load was already on the device before it
was switched on
Take the load off the measuring station
Switch off the measuring station, then
switch it back on again