9. SERVICING
The product must be set up carefully and serviced regularly. Depending on
how frequently the product is used, we recommended servicing at intervals of
3 to 5 years.
CAUTION!
Incorrect measurements as a result of poor servicing
► Have servicing and repairs carried out exclusively by seca Service or
by an authorized service partner.
► You can find service partners in your area at www.seca.com or by
sending an e-mail to service@seca.com.
10.TECHNICAL DATA
10.1 General technical data
The Workflow LED illuminates
in red, an error code
(Er:8:[No.]) is displayed
Errors occurred during the measurement
procedure or data transmission
• Contact your administrator or hospital
technician
Er:8:94 is displayed
An error occurred during the measurement
procedure: IDs were scanned at the wrong
time
• Cancel measurement procedure
• Repeat measurement procedure and
observe proper sequence
• If the problem persists, contact your
administrator or hospital technician
Er:8:95 is displayed
Hardware fault
• Contact your administrator or hospital
technician
• Inform seca Service
The rF menu item is not visible
(seca 797)
The internal interface module of the device
is defective
• Inform seca Service
In the rF menu, only the SYS
menu item is visible
The WiFi function is deactivated • Activate WiFi function
Id:Er is displayed
The ID of the user or the patient has not
been found in the EMR system
• Rescan ID
• If the problem persists, contact your
administrator or hospital technician
Problem Cause Remedy
General technical data
Dimensions
•Depth
•Width
• Height (measuring rod not extended)
576 mm
408 mm
1358 mm
Weight of measuring station Approx. 13.1 kg
Ambient conditions, operation
•Temperature
• Air pressure
•Humidity
+10 °C to +40 °C / +50 °F to 104 °F
700 – 1060 hPa
30 % – 80 %, no condensation
Ambient conditions, storage
•Temperature
• Air pressure
•Humidity
-10 °C to +65 °C / +14 °F to 149 °F
700 – 1060 hPa
0 % – 95 %, no condensation