EasyManua.ls Logo

Seca 787 - Weighing Data

Seca 787
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data • 41
English
17-10-07-652-100/05-2019B
10.2 Weighing data
Ambient conditions, transport
•Temperature
Air pressure
•Humidity
-10 °C to +65 °C / +14 °F to 149 °F
700 – 1060 hPa
0 % – 95 %, no condensation
Power supply (variant-dependent)
Batteries
- Supply voltage
-Battery type
Plug-in power supply unit
- Supply voltage
- Maximum current consumption
- Insulated device, protection class II (EN 60601-1)
9 V
6 x type AA, 1.5 V
12 V
typically 500 mA
Supply voltage 100 V – 240 V
Power supply frequency 50 Hz – 60 Hz
Current consumption
seca 787
seca 797
- WiFi function and USB interface deactivated
- WiFi function and USB interface activated
approx. 55 mA
approx. 55 mA
approx. 250 mA
Maximum runtime in battery operation
seca 787
seca 797
- WiFi function and USB interface deactivated
- WiFi function and USB interface activated
approx. 32 hours
approx. 32 hours
< 5 hours, plug-in power supply unit recommended
USB interface (
seca 797)
Current consumption of the barcode scanner (optional
accessory)
max. 500 mA
Metrology data, measuring rod
Measuring range 1
Measuring range 2
•Graduations
Accuracy
10 cm – 122 cm (4 inch – 48 inch)
122 cm – 230 cm (48 inch – 90.5 inch)
1 mm (1/8 inch)
± 5 mm
Medical device in accordance with Directive 93/42/EEC Class I with measuring function
EN 60601-1: Medical electrical device, Type B
Type of protection IP20
Duty cycle Continuous duty
General technical data
Weighing data
Maximum capacity 250 kg/550 lbs
Minimum capacity 0.2 kg/0.4 lbs
Increments 0.1 kg/0.2 lbs
Tare range up to 250 kg (subtractive)
Accuracy
0 kg to 70 kg (0 lbs to 132 lbs)
70 kg to 250 kg (132 lbs to 551 lbs)
± 100 g (0.22 lbs)
± 0.15 %

Table of Contents

Related product manuals