Technical data • 41
17-10-07-652-100/05-2019B
10.2 Weighing data
Ambient conditions, transport
•Temperature
• Air pressure
•Humidity
-10 °C to +65 °C / +14 °F to 149 °F
700 – 1060 hPa
0 % – 95 %, no condensation
Power supply (variant-dependent)
• Batteries
- Supply voltage
-Battery type
• Plug-in power supply unit
- Supply voltage
- Maximum current consumption
- Insulated device, protection class II (EN 60601-1)
9 V
6 x type AA, 1.5 V
12 V
typically 500 mA
Supply voltage 100 V – 240 V
Power supply frequency 50 Hz – 60 Hz
Current consumption
• seca 787
• seca 797
- WiFi function and USB interface deactivated
- WiFi function and USB interface activated
approx. 55 mA
approx. 55 mA
approx. 250 mA
Maximum runtime in battery operation
• seca 787
• seca 797
- WiFi function and USB interface deactivated
- WiFi function and USB interface activated
approx. 32 hours
approx. 32 hours
< 5 hours, plug-in power supply unit recommended
USB interface (
seca 797)
• Current consumption of the barcode scanner (optional
accessory)
max. 500 mA
Metrology data, measuring rod
• Measuring range 1
• Measuring range 2
•Graduations
• Accuracy
10 cm – 122 cm (4 inch – 48 inch)
122 cm – 230 cm (48 inch – 90.5 inch)
1 mm (1/8 inch)
± 5 mm
Medical device in accordance with Directive 93/42/EEC Class I with measuring function
EN 60601-1: Medical electrical device, Type B
Type of protection IP20
Duty cycle Continuous duty
General technical data
Weighing data
Maximum capacity 250 kg/550 lbs
Minimum capacity 0.2 kg/0.4 lbs
Increments 0.1 kg/0.2 lbs
Tare range up to 250 kg (subtractive)
Accuracy
• 0 kg to 70 kg (0 lbs to 132 lbs)
• 70 kg to 250 kg (132 lbs to 551 lbs)
± 100 g (0.22 lbs)
± 0.15 %