Page 32 Milltronics BW100 - INSTRUCTION MANUAL 7ML19985DJ01
mmmmm
Recalibration
Routine Zero
Run the conveyor empty for several minutes to warm up the belt and insure that it is
empty. Test weights are not used during a zero calibration.
The cause of the increased deviation must be found and rectified. A zero recalibration as
previously described can then be retried.
If the operator deems this deviation to be acceptable, set P377 to 1 to invoke an initial
zero calibration. Further deviation limits are now based on this new initial zero.
Note: E3 is an indication that the mechanical system is errant. The use of P377,
initial zero
, should be used judiciously and only after a thorough mechanical
investigation has been exercised.
1 2
1 2
1 2
1 2
initiate zero calibration
current zero count = ####
zero calculation,
load value displayed
zero deviation calculated
accept deviation
new zero count = ####
return to run mode
1 2
From the
run mode