7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 57
mmmmm
Parameters
P377 Initial Zero 
The initial zero is the reference zero to which all subsequent operator initiated zero 
calibrations are compared in determining whether they have deviated beyond the zero 
limit (P370).
Entry: 
0 = idle 
1 = initial zero 
P388 Initial Span 
The initial span is the reference to which all subsequent span calibrations are compared 
in determining whether they have deviated beyond 12.5% of the initial span. 
Entry: 
0 = idle 
1 = initial span 
Linearization Parameters (P390 - P396) 
P390 Linearization 
Enables or disables the linearization function. 
Entry: 
0 = OFF
1 = ON
P391 Linearizer, Point 1 
Enter the load, in units of P017, for point 1 
P392  Compensation, Point 1 
Enter the calculated compensation, in percent, for compensation point 1 
Note: Refer to 
Recalibration/Initial Zero 
on page 33 for use of this function. 
Note: Refer to R
ecalibration/Initial Span
 on page 36 for use of this function. 
Note: These parameters are used to compensate for non-linear response of the 
weighing system to the BW100. Refer to Recalibration/Linearization for details and 
example on the use of these parameters.