7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 61
mmmmm
Parameters
P648 Totalizer Reset, Internal
Resets the internal totalizers.
Entry:
0 = idle
1 = reset totalizer 2
2 = reset totalizers 1 and 2
End of Totalization parameters.
Communication (P751 - P761)
These parameters are specific to the use of the communication parameters. Refer also to
Communications
, page 47.
P751 Baud Rate
Sets the baud rate for the proprietary bi-polar current loop. This baud rate is not
applicable to communication via Milltronics Comverter.
Entry:
300, 1200, 2400, 4800 or 9600 baud
4800 baud is required for Dolphin communications over the bi-polar current loop.
P760 Communication Mode
Selects the communication mode.
bi-polar current loop: interface with the host device (PLC or computer)
infrared link: communication is made using the Milltronics ComVerter.
maintenance: communication is made via the bi-polar current loop for
RUN mode operations and infrared link for PROGRAM
mode operations
Entry:
1 = bipolar current loop
2 = infrared link
3 = maintenance
P761 Identification Number
Sets the identification number for the unit.
Enter the desired identification number, range 0 - 15.
End of communication parameters.