Page 38 Milltronics BW100 - INSTRUCTION MANUAL 7ML19985DJ01
mmmmm
Recalibration
Factoring 
In order to calculate the value of a new or unknown test weight to the current span, the 
factoring procedure is used. 
Refer to 
Maneuvering
 on page 18 for parameter selection and value change. 
With the conveyor stopped and the belt empty:
Note: For optimum accuracy in the factoring results, a routine zero calibration is 
recommended.
1 2
hang new test weight 
onto belt scale, then 
run the conveyor empty
P359 selected
enter
initiated Factoring,
current test load value
e.g. 76.11 kg / m
dynamic test load value
e.g. 76.03 kg / m
jumps to P017
dynamic test load value
e.g. 76.03 kg / m
return to run mode