7ML19985DJ01 Milltronics BW100 - INSTRUCTION MANUAL Page 65
mmmmm
Parameters
Troubleshooting 
message  diagnosis  action 
E1 (program)  security code required  enter access code into P000 
E2 (run)  totalizer resolution too low  increase value (P631 - P639) 
E3 (run) 
zero out of range  consider an initial zero P377, 
refer to 
Recalibration/Initial 
Zero
 on page 33
E4 (run) 
span out of range  consider an initial span P388, 
refer to 
Recalibration/Initial 
Span 
on
 
page 36
E5 (run) 
parameter not entered  check parameters P005 - P017 
for entry 
E6 (run)  zero calibration required  do a zero calibration 
E7 (run)  span calibration required  do a span calibration 
E8 (program)  parameter value error  check that value is valid 
E9(run) 
auto zero out of range  auto zero has accumulated 
deviation beyond 2% from last 
operator initial zero. If error is 
not caused by material on belt, 
then do an operator initiated 
zero. 
E10 (run) 
rate or span out of range  loading on belt is 300% of rated 
load or greater. Investigate and 
if no mechanical cause, check 
to see if re-rating the design 
rate is required. 
E11 (run) 
speed greater than twice the 
design speed
check design belt speed 
against actual belt speed, 
check speed constant, per-
form speed constant adjust 
(P018) if necessary
E12 (factoring) 
span out of range  test load is either too low or 
beyond 100% of design load 
(P952). Replace test weight 
with a heavier or lighter weight 
and try factoring again.