Page 56 Milltronics BW100 - INSTRUCTION MANUAL 7ML19985DJ01
mmmmm
Parameters
P350 Calibration Security 
This parameter provides additional security to the global lock (P000). 
Entry: 
0 = view parameters, perform zero and span, no reset of totalizer 1 
1 = same as level 0, but cannot perform span 
2 = same as level 0, but cannot perform zero and span 
P359 Factoring 
Factoring is used as a method of calculating the value of the test load (P017) to a new 
physical test weight. 
Entry:
0 = idle 
1 = factor 
P360  Calibration Duration 
Sets the number of belt revolutions to use during a zero or span calibration. 
Enter number of belt revolutions, range 1 - 99.
P367 Direct Zero 
This parameter allows the zero reference count to be viewed or entered directly. 
Direct entry is intended for use when replacing software or hardware and it is not 
convenient to perform an initial zero at that time. 
P368 Direct Span 
This parameter allows the span reference count to be viewed or entered directly. 
Direct entry is intended for use when replacing software or hardware and it is not 
convenient to perform an initial span at that time. 
P370 Zero Limit 
Sets the zero calibration deviation limit from the last initial zero. If the accumulated 
deviation exceeds the limit, the zero calibration is aborted (E3). 
Entry:
0 = +/- 12.5% of initial zero 
1 = +/- 2% of initial zero 
Note: Totalization is halted during the factoring procedure, and resumed only 
upon return to the RUN mode.