Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
General 13
Page 13 of 64
Medical Solutions
1.9 Safety information and protective measures
WARNING
Risk of death, injury or material damage.
Non-compliance can lead to death, injury or material damage.
¹ When performing service work and tests adhere to the
product-specific safety information in the documents.
¹ When performing service work and tests adhere to the
general safety information contained in ARTD Part 2.
¹ Only connect the MOBILETT Plus/Plus E/ Plus M to a
mains power supply outlet (receptacle) that corresponds
to the installation requirements of VDE 0107 or to coun-
try-specific regulations.
¹ Remove or install boards only when the generator is
switched OFF. Adhere to the ESD guidelines e .
¹ Checks and adjustments performed with radiation ON
are identified by the radiation warning symbol x . Dur-
ing these types of adjustments, radiation protective
clothing must be worn.
¹ Perform the protective conductor test according to
ARTD-002.731.17... upon completion of the work and
after all covers have been reattached.
WARNING
When working on the open unit there is a risk of electric shock!
Non-compliance can lead to injury or death.
Observe the following:
¹ Never work with the system open if the capacitor is
charging or has just been charged.
¹ The capacitor bank may still be charged even if the sys-
tem is switched OFF and the mains cable is discon-
nected. There is a risk of life-threatening electric shock.
¹ To avoid danger, refer to the section entitled “Protective
measures when working on the capacitor bank”.