38 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 38 of 64
Medical Solutions
2.4 Service mode
Fig. 20:
Reattach all cover parts with the exception of the upper system cover.
NOTE
Switch S10 on D7 OFF prior to the following inspection.
Entering service mode
Switch S1:1 on board D1 must be switched over (S1:1 to ON) to call up the service pro-
gram. The program version is then displayed:
(depends on the software version installed)
(To exit the service mode, switch S1:1 on board D1 must be switched back to OFF. This
will cause any program that is still running to be shut down.)
Selection of test programs with buttons "kV+/ -"
Several test programs are available. The respective program number appears in the kV
window of the display while the mAs window remains blank. Example of the display for
program no. 6:
kV mAs
vEr x.x = version x.x
kV mAs
P06