Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 31
Page 31 of 64
Medical Solutions
2.2.9 Arm system and single tank
Fig. 11:
SIM Movement of the arm system and the single tank
• The entire arm system including the single tank must not move by itself in any position.
However, it must be easy to move to a new position. If this is not the case, proceed ac-
cording to the section ”Adjusting the friction linings” in the service instructions.
SIM Movement of the single tank fork
SIM Single tank movement
• Check that the single tank moves easily in the fork and stays in any desired position.
• Check that the single tank fork can be easily turned and stays in any desired position.