Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 51
Page 51 of 64
Medical Solutions
2.12 Replacing the collimator lamp
Fig. 31:
PMP Replace every year
Fig. 29:
Fig. 30:
The collimator lamp must be replaced every year.
Remove the cover (1/Fig. 29 / p. 51) by removing the four screws marked ”A”
(Fig. 29 / p. 51). Turn the collimator so that the lamp housing is accessible.
Remove the protective plate (2/Fig. 30 / p. 51) covering the collimator lamp. Loosen
the screws marked ”B” by one turn. Do not remove them. Do not remove nuts ”C”, ”D”
and ”E” in this process (Fig. 30 / p. 51).
Remove the lamp and replace it with a new lamp, (Fig. 31 / p. 51). Do not touch the
glass bulb with your fingers.