36 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 36 of 64
Medical Solutions
PMP Glide bushing on the connecting rod
Lubricate the glide bushing on the connecting rod (Fig. 18 / p. 36).
2.2.12 Arm cable harness
SIM Damage
Fig. 18:
Fig. 19:
Pos. 11 Rotation wear
• Check all the cables for damage. Closely inspect the areas that are frequently subject-
ed to torsion strain (Fig. 19 / p. 36).
• Replace any damaged or broken cables. If you cannot determine any direct cause for
the damage, replace all cable harnesses.
Refer to the section entitled ”Replacement of the support arm cable harness” in the
service instructions.