54 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 54 of 64
Medical Solutions
2.14 Coincidence of radiation and light fields
Fig. 33:
NOTE
If the MOBILETT is equipped with a DAP measuring system, re-
move the DAP ionization chamber first.
Fig. 34:
Set 52 kV and 2.0 mAs on the control and display panel.
X
Release an exposure.
Develop the film and measure the deviation between the light and radiation fields. A de-
viation of ± 1.7% of the SID is acceptable. (Fig. 34 / p. 54)
Insert a film into the 35x35 cm (14"x14") cassette.
Place the cassette on a table or similar surface.
Set the SID to 100 cm (39") (use a tape measure).
Switch the collimator lamp on and align the cassette. Now collimate the light field to
25x25 cm (10"x10").
Mark the four corners of the light field with coins and identify one of the corners with an
additional coin, (Fig. 33 / p. 54)