Siemens AG SPR8-215.831.04.03.02 MOBILETT Plus/ -E/ -M
07.05 CS PS 24
Inspection and maintenance 59
Page 59 of 64
Medical Solutions
2.18 Equivalent leakage current measurement
NOTE
Evaluate the results by comparing the first measured value to the
corresponding value of preceding maintenance procedures or
safety checks. A sudden or unexpected increase in the measured
values may indicate that an error occurred in the primary circuit of
the power supply (damaged insulation, damage caused by water
ingress or humidity, etc.) – even if the limit value of 2 mA is not ex-
ceeded.
SIE Measured value
• Measure the equivalent leakage current according to ARTD-002.731.17... (Safety-
Related Rules for Installation and Maintenance).
The measured values are to be entered and evaluated in the maintenance protocol.
The limit values of 2 mA must not be exceeded.