50 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 50 of 64
Medical Solutions
2.11 DAP measuring system (optional)
Press the test button on the display. The value on the display should be in the range
80-120 µGym
2
(DAP resolution 0.1 µGym
2
) or 8-12 µGym
2
(when using a highly sensi-
tive chamber, i.e. DAP resolution 0.01 µGym
2
). If not, calibrate the system according to
the service instructions.
X
Check the function of the DAP measuring system by releasing an exposure. Check if a
measured value is shown on the display.
SIE Function of the DAP measuring system