50 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 50 of 64
Medical Solutions
2.11 DAP measuring system (optional)
• Press the test button on the display. The value on the display should be in the range
80-120 µGym
2
(DAP resolution 0.1 µGym
2
) or 8-12 µGym
2
(when using a highly sensi-
tive chamber, i.e. DAP resolution 0.01 µGym
2
). If not, calibrate the system according to
the service instructions.
X
• Check the function of the DAP measuring system by releasing an exposure. Check if a
measured value is shown on the display.
SIE Function of the DAP measuring system