32 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 32 of 64
Medical Solutions
2.2.10 Mains cable
Fig. 12:
Pos. 1 Protective conductor
PMP Replace the cable winder every 4 years (units with cable winder only).
The cable winder should be replaced every four years.
SIM Cable winder (for units with cable winder only)
Move the cable up and down while winding it, in order to avoid any slack, until the cable
is completely rolled up. Ensure proper functioning of cable winder and brake.
PMP Replace the mains cable every 2 years.
The mains cable must be replaced every two years.
NOTE
New cable winder 65 08 746 comes with the new mains cable.
Therefore, it is not necessary to replace the mains cable, if the ca-
ble winder has been replaced already.
SIM Damage
Check the mains cable and plug for damage. If necessary, replace them.
PMP Cleaning
Clean the mains cable using a paper towel dampened with a lukewarm solution of wa-
ter and a household cleaning agent.
SIM Protective conductor
Disconnect the power supply plug.