58 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 58 of 64
Medical Solutions
2.17 Protective conductor measurement
SIE Measured value
• Perform the protective conductor test, with all system covers closed, according to
ARTD-002.731.17... (Safety-Related Rules for Installation and Maintenance). The
measured values are to be entered in the maintenance protocol. The protective con-
ductor resistance must not exceed 0.2 ohms.