56 Inspection and maintenance
MOBILETT Plus/ -E/ -M SPR8-215.831.04.03.02 Siemens AG
07.05 CS PS 24
Page 56 of 64
Medical Solutions
2.15 Checking the radiation indicator
Fig. 35:
Pos. 1 Exposure light indicator
X • Release an exposure and check whether an indicator lights up on the control and dis-
play panel and an audible signal sounds at the end of the exposure.
SIE Radiation indicator
SIE Audible signal
• Set 81 kV and 10 mAs.